How high and low do Accuchek Inform II meters read?

How high and low do Accuchek Inform II meters read?

Accu-Chek Inform II Glucose Control Solutions, Roche Diagnostics, Two Levels Level 1 (Low) and Level 2 (High). 1. Store at room temperature at 39-86 °F (4-30° C). Do not freeze.

What is the measuring range of the Inform II glucose meter?

Results were compared to whole blood reference samples, which were analyzed on a Roche/Hitachi 917 analyzer using glucose hexokinase methodology and mathematically converted to IFCC plasma-like reference values. The glucose reference values ranged from 21 to 547 mg/dL, and the hematocrit range tested was 29 to 50%.

What is the testing range of the Accu-Chek Inform II meter?

Highlights of the Inform II System ✓ Sample size – 0.6 microliters and applied to FRONT EDGE TIP of strip ✓ Testing time – 5 seconds ✓ Linear range for meter is 0.6mmol/L-33.3mmol/L ✓ Use of second blood drop for testing ✓ Cleaning site with soap and water ✓ Strip must be in meter BEFORE testing.

When stored properly Accu-Chek Inform II test strips?

Use the test strips at temperatures between 61–95 °F (16–35 °C). Use the test strips between 10–80 % relative humidity. Humidity is the amount of dampness in the air. Store the test strips at temperatures between 39–86 °F (4–30 °C).

What conditions are not appropriate for fingerstick glucose testing?

to perform the test. Patients whose hands are swollen, cold, cyanotic, or edematous are not good candidates for fingerstick (capillary) glucose testing.

In which situations should the inform glucose meter not be used for glucose testing?

patient identification in the glucometer. to perform the test. Patients whose hands are swollen, cold, cyanotic, or edematous are not good candidates for fingerstick (capillary) glucose testing.

Which sample Cannot be used on the Accu-Chek Inform II?

The ACCUCHEK Inform II test strips are for testing fresh capillary, venous, arterial, or neonatal whole blood. Cord blood samples cannot be used.

Do test strips expire?

Using test strips that have expired can cause your results to be inaccurate. If your test strips have expired, throw them away and begin using a new vial of strips. The expiration date is printed on every vial of test strips.

What types of blood Cannot be used with the Accu-Chek Inform II meter and test strips?

What conditions can cause a lower than expected glucose reading?

8 factors that can give an inaccurate blood sugar reading

  • 01/9​8 factors that can give an inaccurate blood sugar reading.
  • 02/9​Expired test strips.
  • 03/9​Extreme temperature can give inaccurate readings.
  • 04/9Sugar, dirt or alcohol on your fingers.
  • 05/9​Wet fingers.
  • 06/9​Dirty meter.
  • 07/9​If you are dehydrated.

What patient conditions can cause the meter to report a falsely decreased value?

Oxygen, acting at location 2, competes with mediator to take electrons from the enzyme. Since strips are generally calibrated for capillary oxygen concentrations, high oxygen values, such as those found in arterial samples or in patients utilizing oxygen, will cause falsely lower values.